All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Abstract

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Author(s): K.Rama Seshaiah, V.Krishna Reddy, Ch.Krishnaiah, VVNKV Prasadaraju, K.Mukkanti, V.Ranga Reddy

A capillary gas chromatographic method using flame ionization detection was developed and validated for the trace analysis (ppm level) of mesityl oxide and diacetone alcohol in pharmaceutical drug substance. The method utilizes a megabore capillary column with bonded and crosslinked polyethylene glycol stationary phase. A dissolve-and-injection approach was adopted for sample introduction in a split mode (1:1). Water is used as sample solvent. A limit of detection of about 1.5 and 2 ppm and limit of quantitation of about 6 and 10 ppm were achieved for the Mesityl oxide and Diacetone alcohol in drug substance samples. The method optimization and validation are also discussed in this paper.


Share this       
Awards Nomination 20+ Million Readerbase

Table of Contents

Google Scholar citation report
Citations : 378

Analytical Chemistry: An Indian Journal received 378 citations as per Google Scholar report

Indexed In

  • CASS
  • Google Scholar
  • Open J Gate
  • China National Knowledge Infrastructure (CNKI)
  • CiteFactor
  • Cosmos IF
  • Electronic Journals Library
  • Directory of Research Journal Indexing (DRJI)
  • Secret Search Engine Labs
  • ICMJE

View More

Flyer