Development and validation of new RP-HPLC method for the estimation of furazolidone in bulk and solid dosage form

Author(s): Dhirender Singh Mittan, Pradeep Kumar, S.P.Venkatesh Prasad, Vikas maurya, Naresh Chandra Joshi, Lalit Kumar

Method is A Rapid, precise, accurate, specific and simple HPLC method developed for estimation of Furazolidone.AHigh performance liquid chromatograph using Phenomenex-Luna RP-18(2), 2504.6mm, 5 mcolumn, with mobile phase having a composition of Methanol: Acetonitrile [90:10 (v/v)] was used. The flow rate of 1.0 ml min-1 and effluent was detected at 259 nm. The retention time of furazolidonewas 2.770minutes. Linearitywas observed in concentration range of 2-18g/ml. Atorvastain was used as internal standard (IS), the retention timewas found to be 2.196minutes. The conc of internal standard used was 8 g/ml in each dilution The mean recoveries were found in the range of 99-100 %. The proposed method was validated for various ICH parameters like linearity, limit of detection, accuracy, precision, ruggedness, robustness, and system suitability.


Share this       
Awards Nomination 20+ Million Readerbase

Table of Contents

Google Scholar citation report
Citations : 378

Analytical Chemistry: An Indian Journal received 378 citations as per Google Scholar report

Indexed In

  • CASS
  • Google Scholar
  • Open J Gate
  • China National Knowledge Infrastructure (CNKI)
  • CiteFactor
  • Cosmos IF
  • Electronic Journals Library
  • Directory of Research Journal Indexing (DRJI)
  • Secret Search Engine Labs
  • ICMJE

View More

Flyer