Application ofmembrane selective electrodes for stability indicating determination of bambuterolHCLin presence of itsmetabolite terbutaline in its pharmaceutical dosage forms and in plasma

Author(s): KhadigaM.Kelani, Wafaa E.Hassan, SheriefM.Eid, AbdallaA.Shalaby4

Two polyvinyl chloride (PVC) membrane sensors for the determination and quantification of Bambuterol HCL (BH) in presence of its acid degradation product (Terbutaline) were described and characterized. The sensors are based on the use of ion association complexes of BH cation with Phosphotungestic acid (PTA) counter anions as ion exchange sites in the PVC matrix(BH-PTA sensor 1) or â- Cyclodextrin (â-CD) to form inclusion complex (BH-â-CD sensor 2). The performance characteristics of the sensors were evaluated according to IUPAC recommendations, reveal a fast, stable and linear response over the concentration range 10-5- 10-2 M for BH.The recoveries for determination of BH by the two proposed selective electrodes were 99.96 ± 0.694, 99.98 ± 0.387, for sensor 1 and sensor 2, respectively. Statistical comparison between the results obtained by this method and the official Pharmacopoeial method of BH was done, and no significant difference was found.


Share this       
Awards Nomination 20+ Million Readerbase

Table of Contents

Google Scholar citation report
Citations : 378

Analytical Chemistry: An Indian Journal received 378 citations as per Google Scholar report

Indexed In

  • CASS
  • Google Scholar
  • Open J Gate
  • China National Knowledge Infrastructure (CNKI)
  • CiteFactor
  • Cosmos IF
  • Electronic Journals Library
  • Directory of Research Journal Indexing (DRJI)
  • Secret Search Engine Labs
  • ICMJE

View More

Flyer